Efficiency Test Of Colimator Shutter At The X Ray Tube In Radiodiagnostic Laboratory Of Poltekkes Jakarta 2 And Two Clinical Hospitals In Jakarta

Main Article Content

Gando Sari
Gita Putri Wahyuni

Abstract

The purpose of this study was to determine the efficiency of the collimator shutter with the presence or absence of leakage from the collector slit of the x-ray equipment. The type of research used is qualitative method using descriptive cross sectional method, conducting collimatory slot testing against x-ray equipment in laboratory and two installation of Radiology field of clinical practice for student of Radiodiagnostic and Radiotherapy of PoltekkesKemenkes Jakarta II. Result, there was a leakage of collimation shutter with marked blackening on x-rayfilm, on 8 (eight) x-ray equipment, there was a full leakage (very significant), and there was a big leak, and there was a slight leakage. Conclusion, in this research there is one x-ray equipment according to the standard reference (golden standard), set by Permenkes No. 1250 / Menkes / SK / XII / 2009 that is SH x-ray equipment in Y hospital.

Article Details

How to Cite
1.
Sari G, Wahyuni G. Efficiency Test Of Colimator Shutter At The X Ray Tube In Radiodiagnostic Laboratory Of Poltekkes Jakarta 2 And Two Clinical Hospitals In Jakarta. SANITAS [Internet]. 1Jun.2017 [cited 22Dec.2024];8(1):16-0. Available from: https://sanitas.e-journal.id/index.php/SANITAS/article/view/20170601-sari
Section
Full Research Article
Author Biographies

Gando Sari, Politeknik Kemenkes Jakarta II

Jurusan Teknik Radiodiagnostik dan Radioterapi Politeknik Kemenkes Jakarta II

Gita Putri Wahyuni, Politeknik Kemenkes Jakarta II

Jurusan Teknik Radiodiagnostik dan Radioterapi Politeknik Kemenkes Jakarta II

References

Carlton R. & AM. Principle of Radiographic Imaging. New York: Delmar Publicer; 2002.

Soekardjo M. Kajian Uji Kesesuaian Pesawat Sinar- X di Instalasi Radiologi. 2015;79–82.

Curry TS. Dowdey JE. & Murry RC. Christensen’s Physics of Diagnostik Radiology. 4th ed. Philadelphia: Lippincott Williams & Wilkins; 1990.

Fauber TL. Radiographic imaging & Exposure. 3rd ed. St Louis: Mosby, 2009

NRCP Report No.99. Quality Assurance For Diagnostic Imaging. Bethesda; 1995.

Peraturan pemerintah no. 63. Keselamatan dan Kesehatan terhadap radiasi pengion. Jakarta; 2000.

International Commision Radiation Protection(IRCP) No.103. Nilai Batas Dosis (NBD). 2007.

Akhadi Mukhlis. Dasar-dasar Proteksi Radiasi. Jakarta: PT Rineka Cipta; 2000.

BATAN. Peraturan Pemerintah Republik Indonesia Nomor 29 Tahun 2008 tentang Perizinan Pemanfaatan Sumber Radiasi Pengion dan Bahan Nuklir. 2008;(2).

BAPETEN. Uji Kesesuaian Pesawat Sinar-X Radiologi Diagnostik dan Intervensional. 2011;

Chesney DN & MO Radiographic Imaging. London: CBS Publicer; 2005.

Keputusan Menteri Kesehatan Republik Indonesia Nomor 1250/MENKES/SK/XII/. Pedoman Kendali Mutu (Quality Control) Peralatan Radiodiagnostik. Jakarta; 2009.